Reference Manual
00809-0100-4664, Rev BA
January 2010
E-33
Rosemount 8712
SIEMENS SENSORS
Connect coil drive and electrode cables as shown in Figure E-31.
Siemens Sitrans F M Mag
3100/5100 Sensors to
Rosemount 8712
Transmitter
Figure E-31. Wiring Diagram for
Siemens Sitrans F M Mag
3100/5100 Sensors and
Rosemount 8712
Table E-29. Siemens Sitrans F M Mag 3100/5100 Sensors Wiring
Connections
Rosemount 8712
Siemens
Sitrans F M Mag 3100/5100 Sensors
1
86
2
85
17
18
82
19
83
82
83
86
85
SIEMENS SITRANS F M MAG
3100/5100 SENSORS
ROSEMOUNT 8712E
TRANSMITTER
This is a pulsed DC magnetic flowmeter.
Do not connect AC power
to the sensor
or to terminals 1 and 2 of the
transmitter
, or replacement of the
electronics board will be necessary.
Summary of Contents for 8712
Page 2: ......
Page 4: ......
Page 26: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 2 16 ...
Page 84: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 4 46 ...
Page 102: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 5 18 ...
Page 114: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 6 12 ...
Page 154: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 D 6 ...
Page 194: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 E 40 ...
Page 208: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 F 14 ...
Page 211: ......