Connect coil drive and electrode cables as shown in Figure E-22.
Figure E-22. Wiring Diagram for
Foxboro Series 2800 and
Rosemount 8712
Table E-20. Foxboro Series 2800/8300 Sensor Wiring Connections
Rosemount 8712
Foxboro
Series 2800/8300 Sensors
1
1
2
2
Chassis Ground
17
Any Shield
18
Black
19
White
ROSEMOUNT 8
TRANSMITTE
FOXBORO SERIES
2800/8300 SENSOR
Outer
Shield
White
White Shield
Black
Black Shield
Inner
Shield
L2
L1
Coil Connections
G
Electrode Connections
This is a pulsed DC magnetic flowmeter.
Do not connect AC power
to the sensor
or to terminals 1 and 2 of the
transmitter
, or replacement of the
electronics board will be necessary.
Summary of Contents for 8712
Page 2: ......
Page 4: ......
Page 26: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 2 16 ...
Page 84: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 4 46 ...
Page 102: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 5 18 ...
Page 114: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 6 12 ...
Page 154: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 D 6 ...
Page 194: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 E 40 ...
Page 208: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 F 14 ...
Page 211: ......