Reference Manual
00809-0100-4664, Rev BA
January 2010
Rosemount 8712
6-4
“Flowrate > 43 ft/sec”
Flow rate is greater than 43 ft/sec
Lower flow velocity, increase pipe diameter
Improper wiring
Check coil drive wiring and sensor coils
Perform sensor test A - Sensor Coil (see Table 6-6 on page 6-9)
“Digital Trim Failure”
(Cycle power to clear
messages, no changes
were made)
The calibrator (8714B/C/D) is not
connected properly
Review calibrator connections
Incorrect calibration number
entered into transmitter
Replace sensor calibration number with 1000005010000001
Calibrator is not set to 30 FPS
Change calibrator setting to 30 FPS
Bad calibrator
Replace calibrator
Table 6-1. Rosemount 8712 Basic Diagnostic Messages
Message
Potential Cause
Corrective Action
Table 6-2. Rosemount 8712 Advanced Diagnostic Messages (Suite 1 - Option Code DA1)
Message
Potential Cause
Corrective Action
Grounding/Wiring Fault
Improper installation of wiring
See “Sensor to Remote Mount Transmitter Connections” on page 2-15
Coil/Electrode shield not
connected
See “Flowtube Sensor Connections” on page 2-17
Improper process grounding
Faulty ground connection
Check wiring for corrosion, moisture in the terminal block, and refer to
“Grounding” on page 5-12
Sensor not full
Verify sensor is full
High Process Noise
Slurry flows - mining/pulp stock
Decrease the flow rate below 10 ft/s (3 m/s)
Complete the possible solutions listed under “Step 2: Process Noise” on
page 6-8
Chemical additives upstream of the
sensor
Move injection point downstream of the sensor, or move the sensor
Complete the possible solutions listed under “Step 2: Process Noise” on
page 6-8
Electrode not compatible with the
process fluid
Refer to the Rosemount Magnetic Flowmeter Material Selection Guide
(00816-0100-3033)
Air in line
Move the sensor to another location in the process line to ensure that it is
full under all conditions
Electrode coating
Use bulletnose electrodes
Downsize sensor to increases flow rate above 3 ft/s (1 m/s)
Periodically clean sensor
Styrofoam or other insulating
particles
Complete the possible solutions listed under “Step 2: Process Noise” on
page 6-8
Consult factory
Low conductivity fluids
(below 10 microsiemens/cm)
Trim electrode and coil wires - refer to “Installation” on page 2-1
Table 6-3. Rosemount 8712 Advanced Diagnostic Messages (Suite 2 - Option Code DA2)
Message
Potential Cause
Corrective Action
8714i Failed
Transmitter Calibration Verification
test failed
Verify pass/fail criteria
Rerun 8714i Meter Verification under no flow conditions
Verify calibration using 8714D Calibration Standard
Perform digital trim
Replace electronics board
Sensor Calibration test failed
Verify pass/fail criteria
Perform sensor test - see Table 6-6 on page 6-9
Sensor Coil Circuit test failed
Verify pass/fail criteria
Perform sensor test - see Table 6-6 on page 6-9
Sensor Electrode Circuit test failed
Verify pass/fail criteria
Perform sensor test - see Table 6-6 on page 6-9
Summary of Contents for 8712
Page 2: ......
Page 4: ......
Page 26: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 2 16 ...
Page 84: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 4 46 ...
Page 102: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 5 18 ...
Page 114: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 6 12 ...
Page 154: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 D 6 ...
Page 194: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 E 40 ...
Page 208: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 F 14 ...
Page 211: ......