![Rosemount 8712 Reference Manual Download Page 173](http://html1.mh-extra.com/html/rosemount/8712/8712_reference-manual_1487460173.webp)
Reference Manual
00809-0100-4664, Rev BA
January 2010
E-19
Rosemount 8712
Model 10D1435 Sensor
(Integral) to
Rosemount 8712
Transmitter
Connect coil drive and electrode cables as shown in Figure E-17.
Figure E-17. Wiring Diagram for
Fischer and Porter Sensor
Model 10D1430 (Integral) and
Rosemount 8712
Table E-15. Fischer and Porter Model 10D1435 Sensor (Integral)
Wiring Connections
Rosemount 8712
Fischer and Porter
Model 10D1435 Sensors (Integral)
1
M1
2
M2
G
17
3
18
1
19
2
Interconnection
Department
ROSEMOUNT 8712
TRANSMITTER
16
18
17
M2
M1
L2/N
L1/L
G
1
2
3
16
17
18
L2/N
L1/L
U1
G
Coil
Connections
Disconnect and
feed through
Magnetic Driver
Assembly
3
U2
G
Electrode
Connections
This is a pulsed DC magnetic flowmeter.
Do not connect AC power
to the sensor
or to terminals 1 and 2 of the
transmitter
, or replacement of the
electronics board will be necessary.
Summary of Contents for 8712
Page 2: ......
Page 4: ......
Page 26: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 2 16 ...
Page 84: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 4 46 ...
Page 102: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 5 18 ...
Page 114: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 6 12 ...
Page 154: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 D 6 ...
Page 194: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 E 40 ...
Page 208: ...Reference Manual 00809 0100 4664 Rev BA January 2010 Rosemount 8712 F 14 ...
Page 211: ......