Rosemount Model 3095 Multivariable Level Controller
A-4
CONNECTIONS AND
HARDWARE
The HART Communicator Model 275 can interface with a transmitter
from the control room, the instrument site, or any wiring termination
point in the loop through the rear connection panel as shown in Figure
A-2. To communicate, connect the HART Communicator in parallel
with the instrument or load resistor. The connections are non-polarized.
Do not make connections to the serial port or NiCad recharger jack in
an explosive atmosphere.
FIGURE A-2. Rear Connection Panel
with Optional NiCad Recharger Pack.
Before connecting the HART Communicator in an explosive
atmosphere, make sure the instruments in the loop are installed
in accordance with intrinsically safe or non-incendive field
wiring practices.
NOTE
The HART Communicator needs a minimum of 250 ohms resistance in
the loop to function properly. The HART Communicator does not
measure loop current directly.
Function
HART Communicator
Fast Key Sequences
Function
HART Communicator
Fast Key Sequences
Software rev
6, 4, 4, 3
Tuning Hi Limit
6, 2, 4, 3, 1
Tag
6, 4, 3, 2
Tuning Lo Limit
6, 2, 4, 3, 2
Target Mode
3
Tuning Rate
6, 2, 4, 8
Time Limit
6, 2, 4, 3, 5
Tuning Reset
6, 2, 4, 7
Tuner
6, 2, 4, 1
Tuning Status
6, 2, 4, 2
Tune Out Max
6, 2, 4, 3, 6
Universal rev
6, 4, 4, 1
Tune Out Min
6, 2, 4, 3, 7
View Status
6, 3, 6
Tuning Bias
6, 2, 4, –
Write protect
6, 4, 3, 7
Tuning Gain
6, 2, 4, 6
Xmtr Height
6, 1, 1, 1, 4
Loop Connection Ports
Serial Port
Optional NiCad
Recharger Jack
2
75-
00
8A
B
Summary of Contents for 3095
Page 2: ......
Page 4: ......
Page 8: ...viii ...
Page 10: ...Rosemount Model 3095 Multivariable Level Controller 1 2 ...
Page 42: ...Rosemount Model 3095 Multivariable Level Controller 3 10 ...
Page 74: ...Rosemount Model 3095 Multivariable Level Controller 5 18 ...
Page 98: ...Rosemount Model 3095 Multivariable Level Controller B 4 ...
Page 101: ...C 3 Approval Drawings FIGURE 3 1 continued 3095 1025A02A ...
Page 102: ...Rosemount Model 3095 Multivariable Level Controller C 4 FIGURE 3 1 continued 3095 1025A03A ...
Page 103: ...C 5 Approval Drawings FIGURE 3 2 Index of I S F M for 3095 3095 1020A01A ...
Page 104: ...Rosemount Model 3095 Multivariable Level Controller C 6 FIGURE 3 2 continued 3095 1020A02A ...
Page 105: ...C 7 Approval Drawings FIGURE 3 2 continued 3095 1020A03A ...
Page 106: ...Rosemount Model 3095 Multivariable Level Controller C 8 FIGURE 3 2 continued 3095 1020A04A ...
Page 107: ...C 9 Approval Drawings FIGURE 3 2 continued 3095 1020A05A ...
Page 108: ...Rosemount Model 3095 Multivariable Level Controller C 10 FIGURE 3 2 continued 3095 1020A06A ...
Page 110: ...Rosemount Model 3095 Multivariable Level Controller C 12 FIGURE 3 3 continued 3095 1024A02A ...
Page 111: ...C 13 Approval Drawings FIGURE 3 3 continued 3095 1024A03A ...
Page 113: ...C 15 Approval Drawings FIGURE 3 4 continued 3095 1021A02A ...
Page 114: ...Rosemount Model 3095 Multivariable Level Controller C 16 FIGURE 3 4 continued 3095 1021A03A ...