DISASSEMBLY
PROCEDURES
Read the following information carefully before disassembling a device.
General information concerning the process sensor body and electrical
housing follow. Figure 2-4 on page 2-6 shows an exploded view of the
Level Controller.
Do not remove the instrument cover in explosive atmospheres when the
circuit is alive.
Process Sensor Body
NOTE
Do not leave devices in service once they have been determined
to be inoperable.
Be aware of the following:
• The process should be isolated from the device and vented before
the device is removed from service for disassembly.
• The process flange can be detached by removing the four flange
bolts and the two alignment screws that secure it.
CAUTION
To prevent damage which may lead to inaccurate measurements, do not
scratch, puncture, or depress the isolating diaphragms.
• Isolating diaphragms can be cleaned with a soft rag, mild
cleaning solution, and clear water rinse.
CAUTION
To prevent damage which may lead to inaccurate measurements, do not
use any chlorine or acid solutions to clean the diaphragms.
• The flange adapters and process flange can be rotated or reversed
for mounting convenience.
• When removing the process flange or flange adapters, visually
inspect the Teflon O-rings. Replace the O-rings if they show any
signs of damage, such as nicks or cuts. If they are undamaged,
they can be reused.
• If the Teflon sensor module O-rings have been replaced, re-torque
the flange bolts after installation to compensate for cold flow.
Summary of Contents for 3095
Page 2: ......
Page 4: ......
Page 8: ...viii ...
Page 10: ...Rosemount Model 3095 Multivariable Level Controller 1 2 ...
Page 42: ...Rosemount Model 3095 Multivariable Level Controller 3 10 ...
Page 74: ...Rosemount Model 3095 Multivariable Level Controller 5 18 ...
Page 98: ...Rosemount Model 3095 Multivariable Level Controller B 4 ...
Page 101: ...C 3 Approval Drawings FIGURE 3 1 continued 3095 1025A02A ...
Page 102: ...Rosemount Model 3095 Multivariable Level Controller C 4 FIGURE 3 1 continued 3095 1025A03A ...
Page 103: ...C 5 Approval Drawings FIGURE 3 2 Index of I S F M for 3095 3095 1020A01A ...
Page 104: ...Rosemount Model 3095 Multivariable Level Controller C 6 FIGURE 3 2 continued 3095 1020A02A ...
Page 105: ...C 7 Approval Drawings FIGURE 3 2 continued 3095 1020A03A ...
Page 106: ...Rosemount Model 3095 Multivariable Level Controller C 8 FIGURE 3 2 continued 3095 1020A04A ...
Page 107: ...C 9 Approval Drawings FIGURE 3 2 continued 3095 1020A05A ...
Page 108: ...Rosemount Model 3095 Multivariable Level Controller C 10 FIGURE 3 2 continued 3095 1020A06A ...
Page 110: ...Rosemount Model 3095 Multivariable Level Controller C 12 FIGURE 3 3 continued 3095 1024A02A ...
Page 111: ...C 13 Approval Drawings FIGURE 3 3 continued 3095 1024A03A ...
Page 113: ...C 15 Approval Drawings FIGURE 3 4 continued 3095 1021A02A ...
Page 114: ...Rosemount Model 3095 Multivariable Level Controller C 16 FIGURE 3 4 continued 3095 1021A03A ...