Trying Out the Instrument
R&S
®
RTE
68
Getting Started 1326.1026.02 ─ 06
6. Insert another mask segment above the positive pulse:
a) Tap the "Append" button under "Mask segments".
b) In the "Region" column of the new mask segment, select "Upper". In this
case, a mask hit is detected if the signal is above the mask limit line.
c) Under "Definition of segment", tap "Insert" twice to insert two points.
d) Enter the x and y-values to define a line beneath which the values of the
positive pulse should remain.
7. Tap the
icon in the result box, and select the "Test Definition" tab.
8. Select channel 1 as the "Source".
9. Define the number of tolerable sample hits in the "Tolerance" field.
A test has failed if the number of sample hits exceeds the limit of "Violation
tolerance" hits.
10. Select the "Event Action / Reset" tab.
11. For the "Stop acquisition" action, select
On violation
. If the violation tolerance
is exceeded, acquisition stops.
The results of the mask test are shown in the "MaskTest" results box. Mask
hits are also indicated as red points in the mask segment in the diagram.
Performing a Mask Test