Trying Out the Instrument
R&S
®
RTE
67
Getting Started 1326.1026.02 ─ 06
The usage of masks tests is also shown in a short video that is available on
the instrument: "Tutorials > Getting Started > Mask Test".
1. To restore the default signal channel settings, press PRESET and AUTOSET.
2. Tap the "Masks" icon on the toolbar.
3. Tap the corner points of the first mask segment on the touchscreen. Select
the corner points of one complete pulse, with a minor offset to the inside. To
finish the mask definition, double-tap the last point.
Tip:
To create an exact rectangle, draw the diagonal of the rectangle on the
screen.
4. To define the further mask test settings, press the MASKS key on the front
panel (in the ANALYSIS area), and select the "Mask Definition" tab.
5. If necessary, correct the mask segment points you defined graphically in the
"Mask Definition" tab. In the "Region" column of the mask segment, "Inner" is
selected. That means, a mask hit is detected if the signal is inside the seg-
ment.
Performing a Mask Test