Command Description
R&S FSH-K1
1157.3564.12 60
E-8
Receiver Mode
The Receiver Mode requires the option R&S FSH-K3.
Command
Parameters
Unit
Comment
CHANNEL <numeric
value>
CHMODE
0, 1
Fixed Channel Mode
CHTABLE <string>
LEVEL
GET command only
MEASTIME <numeric
value>
s
SCANMODE 0,
1
SCANSTART
<numeric value>
Freq Scan Mode
SCANSTOP
<numeric value>
Freq Scan Mode
SCANSTEP
<numeric value>
Freq Scan Mode
FREQSTART
<numeric value>
Channel Scan Mode
FREQSTOP
<numeric value>
Channel Scan Mode
CHMODE (Channel Mode)
This command switches between the channel mode and frequency mode. Use the command
CHTABLE to define a channel table for the channel mode.
Value
Channel Mode
0
Frequency Mode
1
Channel Mode
Example:
set<cr>
0<cr>
chmode,1<cr>
‘selects channel mode
0<cr>
‘responds parameter ok
CHANNEL (Channel Number)
This command defines the channel number. Use CHTABLE command to select a channel table first.
Example:
set<cr>
0<cr>
channel,55<cr>
‘selects channel no. 55
CHTABLE (Channel Table)
This command selects a channel table for the channel mode.
Example:
set<cr>
0<cr>
chtable,FMBand<cr>
‘selects channel table “FMBand”