R&S FSH-K1
Command Description
1157.3564.12 45
E-8
LIMCHKREMOTE (Limit Check Remote Message)
This command enables the remote message (frequency/c level) on limit check and/or on
threshold limit fail. The instrument will send a message (frequency/channel and level) every time the
limit is exceeded.
Value
Limit Check Remote
0
OFF
1
Limit Check – Remote message on
fail
Example:
get<cr>
0<cr>
limchkremote,1<cr>
‘enable remote massage on limit fail
0<cr>
‘responds parameter ok
123.456E6,23.5<cr>
‘limit
exceeded
130.567E6,22.0<cr>
‘limit
exceeded…..
…
0<cr>
‘remote message on limit fail OFF
THRLOW (Lower Threshold Line)
This command defines the lower threshold line (value). The unit of the threshold value is specified with
the Unit parameter. This command is available in Receiver / Scan Mode only.
Example:
set<cr>
0<cr>
thrlow,30<cr>
‘specifies lower threshold line to -30 dBuV
0<cr>
‘responds parameter ok
THRUPP (Upper Threshold Line)
This command defines the upper threshold line (value). The unit of the threshold value is specified with
the Unit parameter. Available in Receiver / Scan Mode only.
Example:
set<cr>
0<cr>
thrupp,70<cr>
‘specifies upper threshold line to 70 dBuV
0<cr>
‘responds parameter ok
THRPASS (Threshold Line Passed Query)
This command returns the threshold line check status. Available in Receiver / Scan Mode only.
Value
Threshold Line Check
0
Unkown
1
Failed
2
Passed
Example:
get<cr>
0<cr>
thrpass<cr>
‘query threshold line check
0<cr>
‘responds parameter ok
2
‘threshold check passed