Command Description
R&S FSH-K1
1157.3564.12 14
E-8
WAIT
This command is used to synchronize with the end of a sweep. After sending the WAIT command the
<ack> acknowledge is holdoff until the sweep is complete.
Example:
cmd<cr>
0<cr>
wait<cr>
‘wait for end of sweep
0<cr>
‘responds parameter ok, sweep complete
STB?
This command returns the instrument status. The status indicates any settings/conditions that causes
measurements to be questionable (response: ‘1’) or not (response ‘0’).
Example:
get<cr>
0<cr>
stb?<cr>
‘query
status
0<cr>
‘responds parameter ok
0<cr>
‘status: o.k.
EXTINPUT (External Input)
This command switches between external trigger input and external reference input on the external
input connector. Default is ‘External Trigger’.
Value
External Input Connector
0
External Trigger
1
External Reference
Example:
set<cr>
0<cr>
extinput,1<cr>
‘enable external reference on input connector
0<cr>
‘responds parameter ok
SAVE
This command saves the current setting and measurement in the specified dataset. If the dataset exists
it will be overwritten. Dataset names are not case sensitive.
Example:
cmd<cr>
0<cr>
save,mydata.001<cr>
‘save settings to dataset “MYDATA.001”
0<cr>
‘responds
parameter
ok
RECALL
This command recalls a stored dataset. Dataset names are not case sensitive.
Example:
cmd<cr>
0<cr>
recall,mydata.001<cr>
‘recall settings from dataset “MYDATA”
0<cr>
‘responds
parameter
ok