Command Description
R&S FSH-K1
1157.3564.12 26
E-8
Trace Settings
Command
Parameters
Unit
Comment
TRACEMODE 0..4
WRAPPHASE 0,1
TRACEDET 0...6
TRACEAVG 2…999
TRACE
<numeric value>,....
GET command only
TRACEBIN
<value>,...
GET command only
TRACETOMEM
CMD command only
CCORRTRACE
<numeric value>,....
GET command only
CCORRTRACEBIN
<value>,...
GET command only
CTRACE
<numeric value>,....
GET command only
CTRACEBIN
<value>,...
GET command only
MATHMODE <numeric
value>
MTRACE
<string>
GET command only
MTRACEBIN
<string>
GET command only
TRACEMODE
This command defines the trace mode.
Value
Trace Mode
0
Clear Write
1
Average
2
Max Hold
3
Min Hold
4
View
Example:
set<cr>
0<cr>
tracemode,2<cr>
‘set trace mode to “Max Hold”
0<cr>
‘responds parameter ok
WRAPPHASE (Phase Wrapping)
This command defines the phase wrapping.
Value
Phase Wrapping
0
Unwrap
1
Wrap
Example:
set<cr>
0<cr>
wrapphase
,0<cr>
‘set wrap phase to unwrap
0<cr>
‘responds parameter ok
Note: This command is available in Tracking Genarator only with measurement modes Smith Chart,
Phase or Vector Magnitude active.