Command Description
R&S FSH-K1
1157.3564.12 18
E-8
Amplitude Settings
Command
Parameters
Unit
Comment
REFLVL
<numeric value>
Uses current unit
REFLVLOFFS <numeric
value>
dB
RANGE 0...20
DYNRANGE 0,
1
UNIT 0...8
RFINPUT 0,
1
PREAMP 0,
1
REFLVL (Reference Level)
This command sets the reference level using the current unit.
Example:
set<cr>
0<cr>
reflvl,-30<cr>
‘set reference level to –30dBm
0<cr>
‘responds parameter ok
REFLVLOFFS (Reference Level Offset)
This command sets the reference level offset in dB.
The new reference level is the current reference level plus the reference level offset.
Example:
set<cr>
0<cr>
reflvloffs,-6<cr>
‘set reference level offset to –6 dB
0<cr>
‘responds parameter ok
RANGE
This command defines the range.
Value
Range
0
10 dB/DIV
1
5 dB/DIV
2
2 dB/DIV
3
1 dB/DIV
4
LIN 0-100%
5
VSWR 1-6 (Tracking Gen, DTF)
6
VSWR 1-2 (Tracking Gen, DTF)
7
VSWR 1-1.5 (Tracking Gen, DTF)
8
VSWR 1-1.1 (Tracking Gen, DTF)
9
Smith Chart
10 0,001
Rho/DIV
11 0,01
Rho/DIV
12 0,1
Rho/DIV
13 1
Rho/DIV