Command Description
R&S FSH-K1
1157.3564.12 42
E-8
TRD1Y (Transducer Y – dBµV/m)
This command selects a Y-direction transducer (dBµV/m only) for the Isotropic Antenna measurement.
To turn a transducer off, use the string ‘NONE’:
Example:
set<cr>
0<cr>
trd1y,ts-emf-y<cr>
‘selects transducer table “TS-EMF-Y”
0<cr>
‘responds parameter ok
Example 2:
set<cr>
0<cr>
trd1y,none<cr>
‘de-activates
any
transducer
0<cr>
‘responds parameter ok
TRD1Z (Transducer Z – dBµV/m)
This command selects a Z-direction transducer (dBµV/m only) for the Isotropic Antenna measurement.
To turn a transducer off, use the string ‘NONE’:
Example:
set<cr>
0<cr>
trd1z,ts-emf-z<cr>
‘selects transducer table “TS-EMF-Z”
0<cr>
‘responds parameter ok
Example 2:
set<cr>
0<cr>
trd1z,none<cr>
‘de-activates
any
transducer
0<cr>
‘responds parameter ok
TRD2 (Transducer – dB )
This command selects a transducer (dB only). To turn a transducer off, use the string ‘NONE’.
Example:
set<cr>
0<cr>
trd1,preamp<cr>
‘selects transducer table “preamp”
0<cr>
‘responds parameter ok
ACCESSORY (Connected Accessory)
This command gets or sets the connected accessory.
Value
Accessory
0
Auto Detect
1
None
2
Bridge FSH-Z2
3
Bridge FSH-Z3
4
Power sensor
5
Isotropic Antenna
Example:
set<cr>
0<cr>
accessory,0<cr>
‘Sets auto detection of accessory”
0<cr>