3 TME400
MANUAL TME400-VCF · EN04 · 2019 March 26th
47
Note
The current parameter settings and meter statuses are lost when re-booting!
They are reset to standard values.
Therefore, prior to booting up, read all parameters of the TME400.
Proceed as follows to re-boot:
•
Switch off the devices
•
Press the "left
◄
" and "right
►
" buttons simultaneously
•
Switch on the voltage again
•
Then, the text "del
A
ll" appears in the display.
•
Release the depressed buttons.
•
Press the calibration button with a thin pencil or small screwdriver.
•
Now the device is booted up and the display shows "Boot".
•
Then, "done" appears in the display and the totalizer status of the main totalizer is
displayed.
Then, re-transmit all device parameters to the TME400 or enter the values from the
test certificated.
Note
The serial interface is set to 38400 Bps, 8N1, Modbus RTU after booting.
These are also the default values of RMGView
EVC
(see chapter 4.4 RMGViewEVC).