Messages
6
SYCON
3000 H
Messages
If the integrated buzzer is activated in the course of a message, you can delete it by pressing the
"HORN" key. The message in the LCD display is only removed when the cause of the message is
remedied or a new analysis is started manually.
You acknowledge an activated relay by pressing the "RELAY" key.
In the case of the display "Analysis stop", you must start a new analysis manually. See the section
"Manual control" on page 5.
Refill indicator
I n d i c a t o r
r e f i l l
P o s s i b l e a n a l y s e s 2 4
H O R N
Indicator must be refilled. On the lower right-
hand side the probable number of possible
analyses is displayed.
As soon as this message is displayed, a certain amount of
indicator is available. As the indicator required per analy-
sis is dependent on the water hardness, the probable
number of analyses cannot be determined exactly in ad-
vance. In calculating the display value, soft water with a
water hardness of 0.0 ppm CaCO
3
is assumed.
After each analysis this value is recalculated with the ac-
tual amount of indicator that still remains.
If there is no indicator when the instrument is
switched on, the display "Indicator lack Analy-
sis stop" appears immediately.
Indicator lack analysis stop
A T T E N T I O N !
I n d i c a t o r l a c k
A n a l y s i s s t o p
H O R N
Automatically no further analyses are started.
Refill indicator.
Error blank value
A T T E N T I O N !
E r r o r b l a n k v a l u e
A n a l y s i s s t o p
H O R N
Automatically no further analyses are started.
Possible causes:
Instrument or sample contaminated
No flush occurred
No water inflow
Calibration necessary
Electrical defect (check connector)
Internal measurement too high
A T T E N T I O N !
I n t r . m e a s u r . t o o h i g h
A n a l y s i s s t o p
H O R N
Automatically no further analyses are started.
Possible causes:
Indicator has not been metered out
No water inflow
Calibration necessary
Electrical defect (check connector)
Internal measurement too low
A T T E N T I O N !
I n t r . m e a s u r e . t o o l o w
A n a l y s i s s t o p
H O R N
Automatically no further analyses are started.
Possible causes:
No water inflow
Calibration necessary
Electrical defect (check connector)
Water hardness over/underrun
A T T E N T I O N !
W a t e r h a r d n e s s
e x c e e d e d
H O R N
According to the programming in program
step 1.5, the overrunning or under running of
the water hardness is displayed.
Example: Hard water = Limit value exceeded.
Permanent signal 1
A T T E N T I O N !
P e r m a n e n t s i g n a l 1
A n a l y s i s s t o p
R E L A Y
Following an overrun or underrun of the water
hardness (see programming step 1.5), the ac-
tivation of the relay "Permanent signal 1" is
displayed.
It is also displayed whether an analysis stop
has occurred or whether analyses continue to
be conducted (see program step 6.3).
Permanent signal 2
A T T E N T I O N !
P e r m a n e n t s i g n a l 2
A n a l y s i s s t o p
R E L A Y
Following an overrun or under run of the wa-
ter hardness (see program step 1.5), the acti-
vation of the relay "Permanent signal 2" is dis-
played. This relay can be activated even in the
case of a fault according to the programming
in program step 6.7. It is also displayed
whether an analysis stop has occurred or
whether analyses continue to be conducted
(see program step 6.5).