RLS Wacon SYCON 3000 H Operating Manual Download Page 8

 

Manual control 

 

 

 

 

SYCON 
3000 H

 

Manual control 

 

The  three  keys  F1

F1

F1

F1,  F2

F2

F2

F2  and  F3

F3

F3

F3  under  the  display  are  characterised  as  softkeys.  These  keys  have  a 

changing function instead of a fixed function. The function that the key has at the moment is shown 
above the key in the inverse colour scheme in the lower display line. In some operating states, you can 
call up further functions for the F3

F3

F3

F3 key with the aid of the KEY

KEY

KEY

KEY button. 

Several  key  functions  are  triggered  after  a  time  delay  to  prohibit  unintended  reactions.  The  running 
time delay is highlighted in the third LCD line. 

 

Analysis start 

 

    A  T  T  E  N  T  I  O  N    !               

A  n  a  l  y  s  i  s    s  t  a  r  t             

        4    s  e  c  o  n  d  s               

S  T  A  R  T   

                           

Press  the  Start  key.  After  4  seconds  a  new 
analysis  sequence  begins  with  a  flush  opera-
tion. 
Attention:  A  new  analysis  can  also  be  started 
when an error message is displayed. 

 

Start extra flush 

    A  T  T  E  N  T  I  O  N    !               

S  t  a  r  t    e  x  t  r  a    f  l  u  s  h       

        4    s  e  c  o  n  d  s               

     

 

     

F  L  U  S  H   

H             

It  is  possible  to  initiate  an  additional  flushing 
process.  In  so  doing,  no  current  values  such 
as e.g. the analysis interval are reset.  
An  additional  flush  is  terminated  again  after 
10  seconds  plus  the  flush  time  entered  in  the 
programme step 1.1.  
By pressing the "END" key a flush operation is 
prematurely aborted.  
Warning! 
Inadequate  flushing  can  lead  to  an  incorrect 
evaluation.  

 

 

Start servicing 

    A  T  T  E  N  T  I  O  N    !               

S  t  a  r  t    s  e  r  v  i  c  e               

        4    s  e  c  o  n  d  s               
     

 

                      S  E  R  V   

Press  the  "SERV"  key.  After  4  seconds  the  ser-
vicing  phase  is  initiated.  You  can  recalibrate 
the  electronics  and  switch  the  metering  pump 
on and off. See also the section Service setting 
1 and 2 on page 21. 
You switch back to normal operation by press-
ing the "END" key. 

 

Test output relays 

    A  T  T  E  N  T  I  O  N    !               

T  e  s  t    o  u  t  p  u  t    r  e  l  a  y  s     

        4    s  e  c  o  n  d  s               
     

 

                      T  E  S  T   

Press the "TEST" key. After 4 seconds the phase Test 
Output relays is initiated. You can switch the 3 out-
put  relays on and off. see also the section "Testing 
the output relays" on page 22..

 

Acknowledge horn  

    A  T  T  E  N  T  I  O  N    !               

T  r  o  u  b  l  e     b  l  a  n  k    v  a  l  u  e   

      A  n  a  l  y  s  i  s    s  t  o  p         

     

 

                   

  H  O  R  N   

If  a  message  appears  in  the  LCD  display  and 
the integrated signal tone generator sounds at 
the  same  time,  the  tone  can  be  deleted  by 
pressing the "HORN" key. 
 
Attention!  The  display  of  the  message  in  the 
LCD  display cannot be deleted. It disappears 
automatically  as  soon  as  the  cause  of  the 
message  is  remedied  or  a  new  analysis  has 
been started.  

 

End 

 

F  l  u  s  h   

 

   

 

                   

L  i  m  i  t   

  2  .  3  p  p  m   

>  =  9  .  0   

  E  N  D                                 

Whenever the key function "END" is displayed, 
the current function can be ended. 

 
 

Acknowledge relays  

    A  T  T  E  N  T  I  O  N    !               

A  c  k  n  o  w  l  e  d  g  e  P  e  r  m  .  r  e  l  1 

      A  n  a  l  y  s  i  s    s  t  o  p         
     

 

      R  E  L  A  Y                 

When  the  key  function  "RELAY"  is  displayed, 
one of the following relays can be deleted: 
 
1. Permanent relay 1 
2. Permanent relay 2 
3. Signal relay 
 
If  the  "RELAY"  key  is  pressed,  the  relay  con-
cerned  is  displayed  and  deleted  after  6  sec-
onds.  
 
If  the  display  "Analysis  stop"  appears,  an 
analysis must be started by hand. 
If programmed accordingly, the start of a new 
analysis can also occur by means of an exter-
nal switch. 

 

Summary of Contents for SYCON 3000 H

Page 1: ...OPERATING MANUAL Analytical Instrument to Measure the Total Hardness of Water SYCON 3000 H...

Page 2: ...urement too low 6 Water hardness over underrun 6 Permanent signal 1 6 Permanent signal 2 6 Input functions 7 Start analysis 7 Stop analysis 7 Reset relay 7 Water meter 7 Output functions 8 Pulse signa...

Page 3: ...function 2 17 Number of the bad messages 17 Activation through a fault 17 Output function Analysis is running 17 Analysis delay 17 Output function Signal relay 18 Fault messages 18 7 Activation of th...

Page 4: ...alve is opened Each measurement begins with an adjustable flushing phase This ensures that the water from the treatment plant is measured and not the water that has remained in the supply line since t...

Page 5: ...isplaying the last result i e Soft water or Hard water blink If both the control lamps Soft water and Hard water are blinking no analysis result is available e g after the instrument has been switched...

Page 6: ...s arranged beneath it e g START and FLUSH Displays during the analyses B l a n c v a l u e 2 s 1 L i m i t 2 3 p p m 9 0 E N D During an analysis the phases Flush Black value Blanc value Titration and...

Page 7: ...F O The functions of the inputs IN1 and IN2 with their current switching states are displayed A horizontal line alongside the designation means Input not active A vertical line alongside the designati...

Page 8: ...Press the SERV key After 4 seconds the ser vicing phase is initiated You can recalibrate the electronics and switch the metering pump on and off See also the section Service setting 1 and 2 on page 2...

Page 9: ...r sample contaminated No flush occurred No water inflow Calibration necessary Electrical defect check connector Internal measurement too high A T T E N T I O N I n t r m e a s u r t o o h i g h A n a...

Page 10: ...analysis starts if the contact of the flow moni tor is activated longer then 10 seconds An analysis also starts if the contact is activated 5 times for 2 seconds e g if water is only taken out for a s...

Page 11: ...6 4 the number of bad messages is programmed after which the sig nal is activated Permanent signal 2 The signal serves to trigger valves horns or signal devices that require a permanent con tact duri...

Page 12: ...low measured value too high measured value too low These faults effect an analysis stop which must be deleted by hand Recorder output RC 2 A recorder at this output permanently registers the displaye...

Page 13: ...seconds until this display appears C H A N G E p r o g r a m m l a n g u a g e P R O G R M L A N G U A E N D 3 Should you wish to change the language of the LCD display press the LANGUA key E N G L I...

Page 14: ...le in step service setting 3 The correction factor needs to be determined empirically with help of a known water sample shift value positive value higher water hardness correction factor higher than 1...

Page 15: ...analysis delay entered A new analysis is only started if the previous one has been completed For this reason an analysis interval can be greater than that programmed in this sec tion Note If a water...

Page 16: ...the desired input function for the input INPUT 1 IN1 with the help of the key Stat Start analysis WaMe Water Meter Stop Stop analysis NoIn No Input Rest Reset relay Input IN 2 S t e p n o 3 2 S t a t...

Page 17: ...a delay time in the range of 1 to 99 sec onds Input function Water meter Water volume 1 between the analyses S t e p n o 4 4 W a t e r v o l 1 1 0 0 c b m Enter the Water volume 1 after which a new a...

Page 18: ...2 Output OUT 2 S t e p n o 5 2 I M P P S 1 P S 2 A N A S I G O U T 2 P e r m a n e s i g n a l 1 Select according to programme step 5 1 the desired output function for the output OUT2 Output OUT 3 S t...

Page 19: ...1 or Permanent signal 2 is so programmed that the ana lyser performs no more analyses before the value entered in programme step 6 2 is reached no further signal can be emitted Output function Perman...

Page 20: ...of the bad messages S t e p n o 6 6 N u m b e r m e s s a g e s 2 You can programme the number of bad messages limit overruns underruns after which the output function is activated Values from 1 to 5...

Page 21: ...tivation of the fault relay RF Refill indicator VH Measured value too high NI No indicator Stop VL Measured value too low BV Blank value too low 7 Activation of the buzzer S t e p n o 7 1 I N B V V H...

Page 22: ...dness of the last analysis is continuously dis played The scale of the display value can be programmed To this end a current initial value of 0 mA or 4 mA must be accorded to the hardness value of 0 d...

Page 23: ...ote the newly entered code number in your documentation Af f f fter ter ter ter a new code number has been entered this programme step cannot be called a new code number has been entered this programm...

Page 24: ...ion The potentiometer required for calibration is now accessible and can be actuated with the help of a screwdriver ATTENTION Replace the cap after calibration S e r v i c e s t e p n o 2 C a l i b r...

Page 25: ...il it reaches a predetermined value W a r t u n g S c h r i t t 3 a l t 0 J N n e u 3 0 s a v e E N D E After the titration has finished the old and new determined shift value is displayed Shift the c...

Page 26: ...eeding Feeding Feeding indicator indicator indicator indicator When commissioning the instrument for the first time the metering pump must be switched on until the indicator flows into the measuring c...

Page 27: ...wall mounting The brackets can be mounted twisted 90 to the left or right For mounting on a plate the in strument can be mounted from the rear side without brackets Use M5 screws The depth of the thre...

Page 28: ...OUT 1 OUT 2 Power in Protectiveearth No Nc C No Nc OUT 3 C No Nc IN 1 IN 2 RC 1 RC 2 L1 Power out n1 l1 Pow er output Pow er input Earth conductor Contact input 1 Recorder output 1 PE PE 4 AmT 500 mA...

Page 29: ...valve with the mains output voltage Inputs The analyser is equipped with 2 identical inputs for floating switching contacts It is only through the programming in the programme steps 3 1 and 3 2 that...

Page 30: ...ms for the regeneration of treatment plants start contact for central control valves with their own pro gramme controller barrels PLC controls stop signal contacts for osmosis plants and fault alarm s...

Page 31: ...Programming Programming Programming Programming 1 5 MAX Activation of the horn if the water hardness is exceeded applies to all outputs 5 1 IMP Generation of a pulse signal 5 4 IMP Activation of the...

Page 32: ...VL Activation on occurrence of the message Internal measurement too low The programming of 1 5 in the case of OUT 1 applies to all outputs Floating fault signal contact at OUT 3 Function Function Func...

Page 33: ...Function Function The recorder serves to log the activities of the instrument Different current values are allocated to the different activities Programming Programming Programming Programming 8 1 4...

Page 34: ...e lever and tighten the hose somewhat 3 Removing the measuring chamber a Remove the metering pump hose b Unfasten the solenoid valve connector and remove it c Loosen and remove the sensor connec tors...

Page 35: ...e sets for maintenance Recommended spare parts for several years 90 142 Maintenance set 01 3x 2 hoses 90161 und 90163 90 362 Measuring chamber cpl without solenoid vale and without intake and outlet p...

Page 36: ...max 3 Pulse signal Permanent signal 1 Permanent signal 2 Analysis runs Signal relay Analogue outputs 2 Output 1 Status display Analysis active Soft water Hard water Refill indicator and Measurement f...

Reviews: