Changing and calling up the programme data
Parameters of the output functions
17
SYCON
3000 H
Output functions: Permanent signal 2
Activation with a limit overrun or underrun
and
in the event of a fault.
Automatic delete function 2
S t e p n o . : 6 . 5
A u t o m a t . d e l e t e 2 Y / N
You can programme whether further analyses are carried out after the activation
of the output relay due to the limit over-/underrun or the faults programmed in
programme step 6.7. If further analyses are conducted, the relay is automati-
cally deactivated again if the result "Water good" occurs in one of the next
analyses.
If no further analyses are conducted, the relay must be deleted by hand or via
the input "Delete relay". Moreover, an analysis must be restarted.
Automatic deletion Yes:
Consecutive
analyses
even
after
a
limit
over/underrun or fault
Automatic deletion No:
Analysis stop in the event of a limit over/underrun
or fault
Number of the bad messages
S t e p n o . : 6 . 6
N u m b e r m e s s a g e s 2 *
#
You can programme the number of bad messages (limit overruns/underruns)
after which the output function is activated. Values from 1 to 5 can be entered.
Activation through a fault
S t e p n o . : 6 . 7
R F | N I | B V | V H | V L |
F a u l t f i l l i n d i A C T I V
#
As opposed to the output function "Permanent signal 1", you can determine with
the output function "Permanent signal 2" which faults cause the relay to be acti-
vated in addition to the limit over/underrun.
Using the "#
#
#
#" key, choose between the display "-" and "|".
"-" = No activation through displayed fault
"|" = Activation through displayed fault
RF
= Refill indicator
VH
= Measured value too high
NI
= No indicator (Stop)
VL
= Measured value too low
BV
= Blank value too low
Output function: Analysis is running
Analysis delay
S t e p n o . : 6 . 8
A n a l y s i s d e l a y 5 s
#
The start of an analysis – opening of the input valve – can be delayed within the
range of 0-999 seconds.
6.9
Output
Bad messages
6.6
*
Permanent
signal 2
Output
Analysis runs
6.2
no
yes
no
?
yes
6.9
6.4
?
Fault messages
6.7
RF
NI
BV
VH
VL
Autom.delete 2
6.5
Yes
No
Delay
6.8
s