Changing and calling up the programme data
Selection of the programmable input functions
13
SYCON
3000 H
3. Selection of the programmable input functions
From the 4 available input functions, maximally 2 can be programmed to the
two inputs IN1
IN1
IN1
IN1 and IN2
IN2
IN2
IN2 of the analyser. ATTENTION: Each input function can
only be used once. If one input is not in use so this input should be pro-
grammed to “NoIn". In programme step 3.3, it is determined whether the inputs
are active with open or closed contact.
Parameters – such as a response delay – must still be entered for the selected
inputs in programme steps 4.1- 4.5.
A description of the inputs can be found in the chapter Input functions on page
7.
Input: IN 1
S t e p n o . : 3 . 1
S t a t S t o p R e s t W a M e N o I n
I N P U T 1 = S t a r t A n a l y s i s
Select the desired input function for the input "INPUT 1 (IN1)" with the help of
the "
" key.
Stat = Start analysis
WaMe
= Water Meter
Stop = Stop analysis
NoIn
= No Input
Rest = Reset relay
Input: IN 2
S t e p n o . : 3 . 2
S t a t S t o p R e s t W a M e N o I n
I N P U T 2 = S t o p A n a l y s i s
Select in accordance with programme step 3.1 the desired input function for the
input "INPUT 2 (IN2)" with the help of the "
" key.
Activation of the input functions
S t e p n o . : 3 . 3
S T A R T | S T O P |
#
Select the way the inputs are activated for the inputs selected in the programme
steps above.
Press the "
" key to select the input functions programmed for the inputs IN1
and IN2 and the "#" key to choose between "|" or "-".
"|" Activation of the selected input function when the contact is closed (NO con-
tact)
"-" Activation of the selected input function when the contact is open (NC con-
tact)
If one input was closed with the function "NoIn", so this possibility is not offered.
4.1
3.3
Stat
Stop
Rest
WatM
Input IN 2
3.2
Activation
2.3
3.1
NoIn
Stat
Stop
Rest
WatM
Input IN 1
NoIn