No. of parameters
1
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
Parameter
Description
Range
Default
1
Format setting
0 .. 12
3
Binary Formats
Format
Data
Order
0
4 Byte (binary)
CRLF
MSB before LSB(=00h)
2
2 Byte (binary)
CRLF
MSB, LSB
4
4 Byte (binary)
CRLF
LSB(=00h) before MSB
6
2 Byte (binary)
CRLF
LSB, MSB
8
4 Byte (binary)
CRLF
MSB before LSB (=Status)
Detect end of answer only by length. CR and LF can be part of the weighing data.
ASCII Formats
Format
Parameter 1
Parameter 2
Parameter 3
Parameter 4
1 & 3
Weight (8)
CRLF
5 & 7
Weight (8)
,
Address (2)
CRLF
9 & 10
Weight (8)
,
Address (2)
,
Status (3)
CRLF
11
Weight (8)
,
Address (2)
,
Extended Status (3)
CRLF
12
Weight (8)
,
Address (2)
,
Extended Status (3)
,
I/O Status (3)
CRLF
Values in brackets signify the number of characters in the fixed length response. The weight format is the sign
(space or minus), followed by 7 digits 0 .. 9 including the decimal point if used.
STATUS Details
Status
Description
Bit
Comment
001
Overload
0
Weight reading out of range overload or
underload
002
Standstill
1
004
Gross
2
008
Range 2 active
3
Only with multi-range or multi-interval
016
I/O 1 state
4
Always 0 in format 12. See Output
Status Details below for Output status.
032
I/O 2 state
5
064
I/O 3 state
6
128
I/O 4 state
7
256
Centre of Zero
8
This status bit is only available in the ex-
tended status contained in formats 11 &
12 only.
Note
The status bits are added together to form the final status. For example a status
of 6 (4 + 2) means the weight reading is a gross value with no motion, range 1,
and all I/O states are inactive.
Output Status Details
129
C500-600-2.0.3
Summary of Contents for C500 series
Page 1: ...C500 Series Digital Indicator Reference Manual ...
Page 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Page 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Page 123: ...118 C500 600 2 0 3 ...
Page 182: ...177 C500 600 2 0 3 ...