Example
Command
Answer
Description
S01;
Select unit 1
ASF?;
9,0 CRLF
Query filtering setting
ASF4,1;
0 CRLF
Changed to a 5 reading average with fine anti-jitter setting
TDD1;
0 CRLF
Save new settings
21.3.6
BDR: Legacy Serial Port Settings
Replaced by the BDX command. Command doesn’t do anything, but is provided for backward compatibility.
General
No. of parameters
4
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
Parameter
Description
Range
Default
1
Baud Rate
1 = 300
2 = 600
3 = 1200
4 = 2400
5 = 4800
6 = 9600
7 = 19200
6
2
Parity
0 = None
1 = Odd
2 = Even
0
3
Data Bits
7, 8
8
4
Stop Bits
1, 2
1
5
Termination Resistors
0 = Off
1 = On
0
Example: Change baud rate settings of unit 1
Command
Answer
Description
S01;
Select unit 1
BDR?;
6,0,8,1,0 CRLF
Query baud rate setting
BDR4,1,7,1,1;
0 CRLF
Settings changed to 2400 baud, odd parity, 7 data bits, 1 stop bit,
termination on.
TDD1;
0 CRLF
Save new settings
Note
If
BDR
is used to change the settings, the reply is sent using the new settings.
21.3.7
BDX: Serial Port Settings
Set the communication parameters, baud rate, parity, etc.
General
No. of parameters
12
Save changes
With
TDD1
Increment Trade Counter
No
Parameter Details
124
C500-600-2.0.3
Summary of Contents for C500 series
Page 1: ...C500 Series Digital Indicator Reference Manual ...
Page 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Page 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Page 123: ...118 C500 600 2 0 3 ...
Page 182: ...177 C500 600 2 0 3 ...