Command
Answer
Description
S01;
Select unit 1
IAD?1;
1,3000,0,1,0,500,1 CRLF
Current settings are:
max1 = 3000,
no decimal point,
e1 = 1,
x10 mode is off,
additive tare is 500,
sense line check is on
IAD1,4000,1,2,0,200,0;
0 CRLF
New settings are:
max1 = 4000,
1 digit after decimal point,
e1 = 2,
x10 mode is off,
additive tare is 200,
sense line check is off
TDD1;
0 CRLF
Save setting
If
IAD?
is issued without the range parameter then the returned data is range 1 for single range setup or range
2 for dual interval or dual range setup. In this way it is possible to query the maximum load without the need
to issue a
WMD?
Command to determine the weighing mode.
21.3.23
ICR: Measurement Rate
Set the fundamental measurement frequency of the instrument.
General
No. of parameters
1
Save changes
With
TDD1
Increment Trade Counter
Yes
Parameter Details
Parameter
Description
Range
Default
1
Measurement Rate in Hertz
10 = 10Hz
12 = 12.5Hz
15 = 15Hz
20 = 20Hz
25 = 25Hz
30 = 30Hz
50 = 50Hz
60 = 60Hz
100 = 100Hz
50
Note
Frequencies other than those listed above will be accepted but the indicator will
operate at the nearest frequency in the table.
Example
Command
Answer
Description
S01;
Select unit 1
ICR?;
50 CRLF
Query current measurement rate
ICR60;
0 CRLF
Change to 60 Hz
TDD1;
0 CRLF
Save setting
136
C500-600-2.0.3
Summary of Contents for C500 series
Page 1: ...C500 Series Digital Indicator Reference Manual ...
Page 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Page 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Page 123: ...118 C500 600 2 0 3 ...
Page 182: ...177 C500 600 2 0 3 ...