Return Format
The query returns the period of the single pulse.
Example
:PULM:PER 1000ms /*Sets the period of the single pulse to 1s (by
default, CH1 is selected).*/
:PULM:PER? /*The query returns 1.*/
3.1.9.4
[:SOURce][:RF]:PULM:POLarity
Syntax
[:SOURce][:RF
<
channel>
]:PULM:POLarity
<
type>
[:SOURce][:RF
<
channel>
]:PULM:POLarity?
Description
Sets the polarity of the pulse modulation for the specified channel.
Queries the polarity of the pulse modulation for the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<type>
Discrete
NORMal|INVerse
NORMal
Remarks
•
NORMal: sets the polarity of the current pulse modulation signal to "Normal".
•
INVerse: sets the polarity of the current pulse modulation signal to "Inverse".
Return Format
The query returns NORM or INV.
Example
:PULM:POL INV /*Sets the polarity of the pulse modulation to INV
(by default, CH1 is selected).*/
:PULM:POL? /*The query returns INV.*/
3.1.9.5
[:SOURce][:RF]:PULM:SOURce
Syntax
[:SOURce][:RF
<
channel>
]:PULM:SOURce
<
type>
[:SOURce][:RF
<
channel>
]:PULM:SOURce?
Command System