Remarks
•
SINGle: selects the "Single" pulse type.
•
TRAin: selects the "Train" pulse type.
This command is invalid when "Ext" modulation source is selected. You can run the
command to set or query the pulse modulation
source.
Return Format
The query returns SING or TRA.
Example
:PULM:MODE SING /*Sets the pulse type of the pulse modulation to
SING (by default, CH1 is selected).*/
:PULM:MODE? /*The query returns SING.*/
3.1.9.2
[:SOURce][:RF]:PULM:OUT:STATe
Syntax
[:SOURce][:RF
<
channel>
]:PULM:OUT:STATe
<
bool>
[:SOURce][:RF
<
channel>
]:PULM:OUT:STATe?
Description
Enables or disables the pulse output for the specified channel.
Queries the on/off status of the pulse output for the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<bool>
Bool
1|ON|0|OFF
0|OFF
Remarks
•
ON|1: enables the pulse output. At this point, the microwave signal generator
can output the pulse signal generated by the internal pulse generator via the
front-panel [PULSE] connector. The output signal is related to the "Mode"
setting.
•
OFF|0: disables the pulse output.
Command System