[:SOURce]:RFALl:OUTPut[:STATe]?
Description
Enables or disables the RF output of all the channels.
Queries the RF output status of all the channels.
Parameter
Name
Type
Range
Default
<bool>
Bool
1|ON|0|OFF
0|OFF
Remarks
•
ON|1: enables the RF output of all the channels.
•
OFF|0: disables the RF output of all the channels.
Return Format
The query returns 1 or 0.
Example
:RFAL:OUTPut ON /*Enables the RF output of all the channels.*/
:RFAL:OUTPut? /*The query returns 1.*/
3.1.11
[:SOURce][:RF]:MODulation Commands
3.1.11.1
[:SOURce][:RF]:MODulation:STATe
Syntax
[:SOURce][:RF
<
channel>
]:MODulation:STATe
<
bool>
[:SOURce][:RF
<
channel>
]:MODulation:STATe?
Description
Enables or disables the modulation output for the specified channel.
Queries the on/off status of the modulation output for the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<bool>
Bool
1|ON|0|OFF
0|OFF
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
71