Return Format
N/A
Example
N/A
3.3
:UNIT Command
The :UNIT command is used to set the output and displayed units for the instrument.
3.3.1
:UNIT[:RF]:POWer
Syntax
:UNIT[:RF
<
channel>
]:POWer
<
unit>
:UNIT[:RF
<
channel>
]:POWer?
Description
Sets the unit for the output and displayed amplitude of the specified channel.
Queries the unit for the output and displayed amplitude of the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<unit>
Discrete
dBm|dBmV|dBuV|V|W
dBm
Remarks
N/A
Return Format
The query returns dBm, dBmV, dBuV, V, or W.
Example
:UNIT:POW V /*Sets the unit for the output and displayed amplitude
of CH1 to V (by default, CH1 is selected).*/
:UNIT:POW? /*The query returns V.*/
3.4
:SYSTem Commands
The :SYSTem commands are used to set a series of parameters relating to the system
and the settings of these parameters do not affect the output signal of the microwave
signal generator.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
79