Return Format
The query returns the stop frequency of the step sweep.
Example
:SWE:STEP:STOP:FREQ 4MHz /*Sets the stop frequency of the step
sweep to 4MHz (by default, CH1 is selected).*/
:SWE:STEP:STOP:FREQ? /*The query returns 4000000.*/
3.1.5.22
[:SOURce][:RF]:SWEep:STEP:STOP:LEVel
Syntax
[:SOURce][:RF
<
channel>
]:SWEep:STEP:STOP:LEVel
<
level>[<unit>]
[:SOURce][:RF
<
channel>
]:SWEep:STEP:STOP:LEVel?
Description
Sets the stop level of the step sweep for the specified channel.
Queries the stop level of the step sweep for the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<level>
Real
-30 dBm to 25 dBm
-20 dBm
<unit>
Discrete
dBm|V|W|dBmV|dBuV
dBm
Remarks
•
When <level> is set in "Number" form (for example, 2), the default unit is dBm.
When set in " Unit" form (for example, 2dBm), the start level
displayed in the interface of the microwave signal generator is related to the
setting of the level unit.
-
When the level unit is "dBm", 2.00dBm is displayed.
-
When the level unit is "dBmV", 48.99dBmV is displayed.
-
When the level unit is "dBuV", 108.99dBuV is displayed.
-
When the level unit is "Volts", 281.50mV is displayed.
-
When the level unit is "Watts", 1.58mW is displayed.
•
The default unit of the return value is dBm.