Remarks
When editing the sweep list, you can use this command to add one line of sweep
values. At this point, one line is added to the sweep list.
•
<freq>: sets the frequency of the sweep point.
•
<level>: sets the amplitude at the frequency point.
•
<time>: sets the duration of a sweep step.
Return Format
N/A
Example
:SWEep:RF2:LIST:ADDList 1M,10,1s /*Adds one line of sweep values
(frequency 1MHz, amplitude 10dBm, dwell time 1s) to the sweep list
for CH2.*/
3.1.5.4
[:SOURce][:RF]:SWEep:LIST:CPOint
Syntax
[:SOURce][:RF
<
channel>
]:SWEep:LIST:CPOint?
Description
Queries the number of points in the current sweep list for the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
Remarks
You can use the
[:SOURce][:RF]:SWEep:LIST:DELete
command to delete one line of
sweep values or use the
[:SOURce][:RF]:SWEep:LIST:ADDList
command to add one
line of sweep values.
Return Format
The query returns the total number of sweep points in the sweep list. For example, 5.
Example
N/A
Command System
20