When the frequency offset is 0 Hz, the frequency displayed in the interface
(namely the setting frequency) is equal to the actual output frequency.
•
When the frequency offset is not 0 Hz, the output frequency is determined by
the setting frequency and frequency offset. The three parameters satisfy the
equation: setting frequency (display frequency) = output fre frequency
offset. You can query or set the RF frequency by using the
command.
Return Format
The query returns the RF frequency offset of the specified channel.
Example
:RF2:FREQ:OFFS 20kHz /*Sets the RF frequency offset of CH2 to
20kHz*/
:RF2:FREQ:OFFS? /*The query returns 20000Hz.*/
3.1.2
[:SOURce][:RF]:LEVel Commands
3.1.2.1
[:SOURce][:RF]:LEVel
Syntax
[:SOURce][:RF
<
channel>
]:LEVel
<
level>[<unit>]
[:SOURce][:RF
<
channel>
]:LEVel?
Description
Sets the RF output amplitude of the specified channel.
Queries the RF output amplitude of the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<level>
Real
-30 dBm to +25 dBm
-30 dBm
<unit>
Discrete
dBm|V|W|dBmV|dBuV
dBm
Command System