If the current sweep direction is "Down", the instrument will stop the current
sweep and sweep from the stop frequency or stop level after sending this
command.
Return Format
N/A
Example
N/A
3.1.5.15
[:SOURce][:RF]:SWEep:POINt:TRIGger:TYPE
Syntax
[:SOURce][:RF
<
channel>
]:SWEep:POINt:TRIGger:TYPE
<
type>
[:SOURce][:RF
<
channel>
]:SWEep:POINt:TRIGger:TYPE?
Description
Sets the point trigger type of the sweep for the specified channel.
Queries the point trigger type of the sweep for the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<type>
Discrete
AUTO|KEY|BUS|EXT
AUTO
Remarks
•
AUTO: selects Auto trigger.
If the sweep mode is "Continuous", select one sweep type and the instrument
sweeps each sweep point continuously within a sweep period. If the sweep
mode is "Single", you need to send the
command first to meet the single sweep condition, and then it will launch one
sweep period and then stops.
•
KEY: selects Key trigger.
If the sweep mode is "Continuous", every time you click or tap Key, the
instrument starts to sweep one point. If the sweep mode is "Single", you need
to send the
command first to meet the single
sweep condition. At this point, every time you click or tap Key, the instrument
sweeps one point and then stops after completing one sweep period.
Command System
28