Example
:RF2:AM:SOUR EXT /*Sets the AM modulation source type of CH2 to
EXT./
:RF2:AM:SOUR? /*The query returns EXT.*/
3.1.6.6
[:SOURce][:RF]:AM:EQUal:LEVel
Syntax
[:SOURce][:RF
<
channel>
]:LEVel:AM:EQUal:LEVel
<
value>
[:SOURce][:RF
<
channel>
]:LEVel:AM:EQUal:LEVel?
Description
Sets the on/off status of the RF equal level output of the specified channel.
Queries the on/off status of the RF equal level output of the specified channel.
Parameter
Name
Type
Range
Default
<channel>
Discrete
1|2|3|4|5|6|7|8
1
<value>
Bool
1|ON|0|OFF
0|OFF
Remarks
N/A
Return Format
The query returns 1 or 0.
Example
:AM:EQUal:LEVel ON /*Enables the RF equal level output of CH1 (by
default, CH1 is selected).*/
:AM:EQUal:LEVel? /*The query returns 1.*/
3.1.6.7
[:SOURce]:AM:EXT:COUP
Syntax
[:SOURce]:AM:EXT:COUP
<
type>
[:SOURce]:AM:EXT:COUP?
Description
Sets the AM external coupling mode.
Queries the AM external coupling mode.
Command System