■Patient environment
The patient environment represents a space where there is a possibility of direct contact
between the patient or the operator and third person.
When another type of device is used in the patient environment, use a device that
complies with IEC 60601-1. If the devices that do not comply with IEC 60601-1 are used,
it is necessary to use an isolating transformer to power the device or to connect the
devices to additional protective grounding.
Radius of 1.5m
1.5m
1.5m
2.5m