30
1-2 Instrument Classification
<Classification as per 93/42EEC (MDD)>
Class
Ⅱ
a
<Protection type against electric shock>
Class
Ⅰ
This instrument is classified as equipment whose protection against electric shock
does not rely on basic insulation only, but which includes an additional safety
precaution, in that means are provided for the connection of the equipment to a
protective earth conductor in the fixed wiring of the installation in such a way that
accessible metal parts cannot become live in the event of a failure of the basic
insulation. Always use the power cord shipped with the instrument. It must be
connected to a receptacle outlet with a ground.
<Degree of protection against electric shock>
Type B Applied Part
The forehead rest of this instrument is a type B attachment. This instrument is
equipped with a reliable safeguard against electric shock, as verified through
testing in accordance with the applicable standard for patient leakage current and
measurement current.
<Protection against water and particle ingression>
IP20
This instrument is not protected against the ingression of water and other liquids.
The instrument should not be exposed to liquids.
<Degree of safety for use in a flammable environment>
Retinomax 3 is not designed to be used in a flammable environment. Do not use it
in a flammable environment.
<Disinfection method approved by the manufacture>
If necessary, clean the forehead rest with soft cloth soaked with rubbing alcohol.
Cleaning conducted by a process other than the one described above may cause
deterioration or damage to the equipment.
<Operation mode>
Continuous operation
<Laser classification>
EN60825-1.2:2001 Class 1 LED product
Wavelength
855 nm to 870 nm
Radiant power
111