■Automatic Data Backup Function
The automatic data backup function allows you to redisplay and print out the data
obtained before the power shutdown in the following circumstances by
pressing
key after turning on the power.
•
The power went off while carrying the main unit.
•
The power was turned off by mistake.
•
The power went off during data output such as printing or external delivery.
1. Turn on the power, and press
. The LAST DATA screen appears, and the
data effective immediately before the power shutdown is displayed, printed out
or externally delivered.
•
•
If you access the menu screen, the data effective immediately before the
power shutdown will be deleted.
•
The automatic data backup function does not work if the
retroillumination mode
“5
-
6 Retro illumination Mode”
was used
immediately before the power shutdown.
•
If power goes down during measurement, the data effective
immediately before the power shutdown is not outputted if the
measured values have not been obtained.
•
The output time indicates the time when the power was turned on.
Note that it is not the time when power went off.
•
If the battery pack runs down, the automatic data backup function does
not work.
LAST DATA screen
Time when the
key is pressed after the
power goes off