6
•
Never short-circuit the battery charging contacts of the main
unit, station, or printer. If the contacts are short-circuited, the
fuse of the station or printer will blow.
•
Do not expose the connectors or battery charging contacts to
your body or metal objects. If a contact is dirty, turn the power
off and wipe the contact off with a dry, soft cloth.
•
This product is precision optical equipment, containing a large
number of electrical components. Handle with care, avoiding
exposure to physical shock.
•
Do not drop or bump the instrument. When handling the main
unit, always use the supplied strap. Do not swing the main
unit holding only the strap.
•
If the instrument ceases to function, pull the power cord from
the receptacle outlet, and without accessing the inside of the
instrument, contact your dealer.
•
Use the designated print roll. Use of an undesignated brand
can lead to failure.
•
The operator must not touch the external instrument and the
external monitor connected with this equipment and the
patient simultaneously. Electrical shock can result.
•
When the device is return for maintenance or repair, clean the
surface of the device (especially, forehead rest) with a cloth
dampened with rubbing alcohol for disinfections.
•
Accessory equipment connected to the analog and digital
interfaces must be certified according to the appropriate
national standards.
Furthermore, all configurations shall comply with the IEC
60601-1. Anyone who connects additional equipment to the
signal input part or signal output part is responsible for
making sure that the system complies with the requirements of
the IEC 60601-1.
If in doubt, consult the technical service department or your
local representative.
CAUTION 6. Maintenance/Storage
•
The ambient conditions of transportation: Temperature -20°C to
50°C, relative humidity 10% to 95% (no condensation),
atmospheric pressure 500-1060 hPa
•
The ambient conditions of
l
in a packaged condition:
Temperature -10°C to 40°C, relative humidity 10% to 95%(no
condensation), atmospheric pressure 700-1060 hPa
This instrumentdoes not meet the temperature requirementsof
ISO 15004-1 for storage. Do not store this instrument in
conditions where the temperature may rise above 40°C or fall
below -10°C.
•
Select a storage area exposed to limited amounts of dust, free
of vibration and shock.