30. Reinstall the first scanner and fix the front guides (1 screw each).
– First Scanner Position Adjustment –
31. Set four positioning pins [A] to the first and second scanners as shown.
32. Fix the scanner wires to the first scanner using the front and rear wire
clamps as shown (2 screws each).
33. Remove the scanner positioning pins [A] and slide the scanner back and
forth several times manually. Then, confirm the scanner position by
resetting the positioning pins.
NOTE: If they cannot be set smoothly, repeat steps 27 to 33.
34. Reassemble the machine.
35. Perform the scanner free run by the SP test mode and check the scanner
movement.
36. Make copies and confirm the copy quality, adjust the magnification and
focus by SP mode if necessary.
[A]
[A]
A171R551.img
31 July 1995
OPTICS
5-19