Auto Max
This screen allows you to modify the Auto maximum pressure
setting. You can adjust this setting from the Auto minimum
pressure setting to 20 cm H2O. This screen only displays if
Auto-CPAP mode is enabled or if the Auto-Trial feature is
available and enabled.
cmH
2
O
Pressure
This screen allows you to adjust the CPAP pressure, or the
CPAP-Check mode starting pressure. If Auto-Trial mode was
used, you can choose the 90% pressure setting determined
from the Auto-Trial mode, or you can adjust this setting from 4
to 20 cm H2O. If the Auto-Trial mode was not used, this
screen allows you to only adjust the pressure setting from 4 to
20 cm H2O.
IPAP
IPAP
This screen allows you to modify the IPAP setting. The initial
default setting is 20 cm H2O. You can adjust the setting from
the EPAP setting to 25 cm H2O. This screen only displays if
Bi-level mode is enabled.
EPAP
EPAP
This screen allows you to modify the EPAP setting. The initial
default setting is 4 cm H2O. You can adjust the setting from 4
cm H2O to the IPAP setting. This screen only displays if Bi-
level mode is enabled.
IPAP Max
This screen allows you to modify the Maximum IPAP setting.
The setting you specify here will be the maximum level of
pressure applied during the inspiratory breath phase. You may
adjust the setting from the Minimum EPAP setting to 25 cm
H2O. This screen only displays if Auto Bi-level mode is
enabled.
EPAP Max
This screen allows you to modify the Minimum EPAP setting.
The setting specified here will be the minimum level of
pressure applied during the expiratory breath phase. You may
adjust the setting from 4 cm H2O to the Maximum IPAP
setting. This screen only displays if Auto Bi-level mode is
enabled.
PS Min
This screen allows you to modify the Minimum Pressure
Support setting. This setting is the minimum difference that is
permitted between IPAP and EPAP while Auto
Bi-level therapy mode is active. You may adjust the setting
from 0 cm H2O to the Maximum Pressure Support setting.
This screen only displays if Auto Bi-level mode is enabled.
PS Max
This screen allows you to modify the Maximum Pressure
Support setting. This setting is the maximum difference that is
permitted between IPAP and EPAP while Auto Bi-level therapy
mode is active. You may adjust the setting from 0 cm H2O to
the minimum value of either 8 cm H2O, or the difference
between Max IPAP and Min EPAP. This screen only displays if
Auto Bi-level mode is enabled.
I
CON
T
EXT
D
ESCRIPTION
Summary of Contents for DreamStation
Page 1: ...DreamStation Service Technical Reference Manual ...
Page 5: ...1117539 VER 07 This page intentionally blank ...
Page 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Page 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Page 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Page 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Page 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Page 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...