P
AGE
4-9
1117539, V
ER
. 07
2.
If applicable, place a clean, reusable blue pollen filter (1) on top of a new, optional disposable light-
blue ultra-fine filter (2) and firmly snap them together.
F
IGURE
4-14: P
OLLEN
AND
U
LTRA
-F
INE
F
ILTERS
3.
Place the new filter assembly back in the side of the therapy device. Swing the door closed.
F
IGURE
4-15: I
NSTALLING
THE
F
ILTER
4.9 S
TARTING
THE
D
EVICE
1.
Ensure power is supplied to the device. The first screen to display will be the Philips Respironics
logo, followed by the device model screen, and then the Home screen.
F
IGURE
4-16: H
OME
S
CREEN
The first time the device is powered on, a pop-up will prompt you to set the time on the device. The default
setting is Greenwich Mean Time, but you may adjust the time in 30 minute increments to match your local time
zone. If you choose to skip this initial time setting, the time can always be adjusted under the “My Setup” menu.
Note:
This time setting is not displayed as a clock function on the device. It is only used to align
therapy data for Provider’s data reports.
Summary of Contents for DreamStation
Page 1: ...DreamStation Service Technical Reference Manual ...
Page 5: ...1117539 VER 07 This page intentionally blank ...
Page 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Page 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Page 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Page 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Page 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Page 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...