E-201
ERR_RTOS_TIMEOUT
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-202
ERR_RTOS_ONE_SHOT_
TASK_NOT_STOPPED
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-210
ERR_PERF_CHECK_MS
G_TO
STOP
PCA
1. Replace PCA and test.
E-211
ERR_PERF_CHECK_POS
T_FAIL
STOP
PCA
1. Replace PCA and test.
E-220
ERR_MOTOR_ID_TIMEO
UT
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-221
ERR_MOTOR_SHUNT_S
AFETY_TIMEOUT
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-222
ERR_MOTOR_DIV_ZERO
CONTINUE
N/A
1 Clear error log and retest.
E-230
ERR_HUMID_PLATE_CY
CLE_OVERRUN
REBOOT
PCA
1. Clear error log and test.
2. Test humidifier
separately.
E-231
ERR_HUMID_PLATE_NO
_MEASUREMENT
REBOOT
1. Heater Plate
2. PCA
1. Clear error log and test.
2. Test humidifier
separately.
E-240
ERR_PACKET_CREATIO
N
CONTINUE
N/A
1. Clear error log and test.
E-241
ERR_INVALID_COMPLIA
NCE_SIZE
CONTINUE
N/A
1. Clear error log and retest.
E-250
ERR_BAROMETRIC_CO
MM
CONTINUE
PCA
1. Clear error log and test.
2. If the test fails or the error
reoccurs, replace the PCA
and retest.
C
ODE
E
RROR
N
AME
E
RROR
L
EVEL
F
AILED
C
OMPONENT
A
CTIONS
Summary of Contents for DreamStation
Page 1: ...DreamStation Service Technical Reference Manual ...
Page 5: ...1117539 VER 07 This page intentionally blank ...
Page 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Page 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Page 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Page 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Page 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Page 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...