![Respironics DreamStation Service & Technical Reference Manual Download Page 120](http://html.mh-extra.com/html/respironics/dreamstation/dreamstation_service-and-technical-reference-manual_1445207120.webp)
P
AGE
7-17
1117539, V
ER
. 07
R
EMOVING
/
INSTALLING
THE
B
OTTOM
/F
LIP
L
ID
A
SSEMBLY
:
1.
Remove the Tank Assembly, Flip Lid Seal, Dry Box Seal, and the Back Panel Assembly (refer to
previous corresponding sections).
2.
Install Back Panel Assembly, Flip Lid Seal, Dry Box Seal and Tank Assembly onto the new Bot-
tom/Flip Lid Assembly as instructed in the previous sections.
3.
Place a new warning label and Serial/Model number label on the bottom of the Humidifier (refer to
Section 6.2
for creating the Serial/Model number label.
F
IGURE
7-25: W
ARNING
L
ABEL
AND
S
ERIAL
/M
ODEL
N
UMBER
L
ABEL
P
LACEMENT
4.
Route the heated tube wire harness through the Flip Lid opening as shown below.
F
IGURE
7-26: H
EATED
T
UBE
W
IRE
H
ARNESS
R
OUTED
T
HROUGH
F
LIP
L
ID
Summary of Contents for DreamStation
Page 1: ...DreamStation Service Technical Reference Manual ...
Page 5: ...1117539 VER 07 This page intentionally blank ...
Page 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Page 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Page 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Page 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Page 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Page 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...