Incorrect
Power
Supply
Notification
Indicates an
incompatible power
supply is attached.
Incompatible
power supply,
or power cord is
not fully inserted
into device’s
power inlet.
Confirm power
cord is full
inserted into
device’s power
inlet. Confirm a
compatible
Philips
Respironics
power supply is
attached.
Switch to
compatible
power supply if
needed.
Low Voltage
Notification
Low voltage.
Incompatible
power supply is
attached.
Confirm a
compatible
Philips
Respironics
power supply is
attached.
Switch to
compatible
power supply if
needed. If
battery is being
used, ensure
battery is
adequately
charged.
Automatic Off
Status
Displayed when
therapy ends due to
automatic off
function.
The mask has
been removed.
n/a
Inlet blocked.
Check filter.
Notification
Blocked airway
Blockage at
device
inlet.
Check device air
inlet is not
obstructed.
Check air
filter(s) are
installed properly
and are clean;
replace if
needed.
Low Leak:
Check Mask
and Tube
Notification
Blocked airway
Blockage at
tube or mask.
n/a
A
LERT
I
CON
T
YPE
D
ESCRIPTION
P
OSSIBLE
C
AUSE
A
CTION
Summary of Contents for DreamStation
Page 1: ...DreamStation Service Technical Reference Manual ...
Page 5: ...1117539 VER 07 This page intentionally blank ...
Page 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Page 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Page 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Page 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Page 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Page 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...