To verify the pressure, complete the following steps:
1.
Install the blue pollen filter into the device.
2.
With the device unplugged, connect the system as illustrated in the diagram.
3.
Turn the manometer on. If it does not display a reading of zero, adjust the manometer to calibrate
it. If the manometer has variable settings for devices, set it to cm H2O.
4.
Supply power to the device then place the device in provider mode.
5.
Set the therapy parameters according to the patient specific data.
6.
Set the device to the specific pressure value for the patient.
7.
Verify that the pressure setting matches the pressure displayed on the manometer. If the pressure
setting does not match the measured value for the device, contact Philips Respironics or an autho-
rized service center to have the device serviced.
8.
Set up the remaining parameters and exit provider mode. The unit is ready for patient use.
5.3 S
ERVICE
C
ENTER
T
OOLS
S
UITE
The Service Center Tools Suite will provide you the necessary tools to view the device’s error/event log, along
with additional functions necessary to service the device. To download the software you must log onto
my.respironics.com. If you do not have an account, click on the “Sign Up” link to register for an account.
F
IGURE
5-1: M
Y
.
RESPIRONICS
.
COM
Login Here
http://my.respironics.com
Service Software and
Documentation Link
Summary of Contents for DreamStation
Page 1: ...DreamStation Service Technical Reference Manual ...
Page 5: ...1117539 VER 07 This page intentionally blank ...
Page 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Page 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Page 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Page 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Page 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Page 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...