P
AGE
7-4
1117539, V
ER
. 07
T
O
I
NSTALL
THE
S
EALS
:
1.
Press the Seals into the place as shown in Figures 7-2 and 7-3.
2.
Press the Tank Top Seal onto the Patient Outlet Swivel Clip.
Note:
The Seal only aligns one way. Be sure to press the Seal the whole way around to ensure it is seated
properly.
7.1.3 R
EPLACING
THE
D
RY
B
OX
A
SSEMBLY
T
O
REMOVE
THE
D
RY
B
OX
A
SSEMBLY
:
1.
Remove the Water Tank and Dry Box Seal (refer to previous corresponding sections).
2.
Pull back on the Dry Box Latch and pull the Dry Box out of the Bottom Assembly.
F
IGURE
7-4: D
RY
B
OX
A
SSEMBLY
T
O
I
NSTALL
THE
D
RY
B
OX
A
SSEMBLY
:
1.
Position the Dry Box into the Bottom Assembly.
2.
Snap the Dry Box into place and verify it is fully seated.
3.
Assemble the remainder of the device as instructed in the previous sections.
Kit
Included in Kit
Tools Required
Dry Box Assembly
•
Dry Box Assembly
•
None
Pull back on latch
to remove
Dry Box Assembly
Summary of Contents for DreamStation
Page 1: ...DreamStation Service Technical Reference Manual ...
Page 5: ...1117539 VER 07 This page intentionally blank ...
Page 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Page 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Page 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Page 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Page 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Page 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...