6
•
pneumothorax
•
pathologically low blood pressure
•
dehydration
•
cerebrospinal fluid leak, recent cranial surgery, or trauma.
W
ARNINGS
•
The entire manual should be read before using the
S8 A
UTO
S
ET
V
ANTAGE
.
•
Advice contained in this manual should not supersede instructions given by the
prescribing physician.
•
A patient should not connect a device to the data communication port unless
instructed to do so by their health care provider or physician. Only ResMed
products are designed to be connected to the data communication port.
Connecting other devices could result in injury, or damage to the
S8 A
UTO
S
ET
V
ANTAGE
device.
•
The
S8 A
UTO
S
ET
V
ANTAGE
should only be used with masks (and connectors)
*
recommended by ResMed, or by a physician or respiratory therapist. A mask
should not be used unless the
S8 A
UTO
S
ET
V
ANTAGE
device is turned on and
operating properly. The vent hole or holes associated with the mask should never
be blocked.
Explanation:
The
S8 A
UTO
S
ET
V
ANTAGE
is intended to be used with special
masks (or connectors)
*
which have vent holes to allow continuous flow of air out
of the mask.
When the device is turned on and functioning properly, new air from the device
flushes the exhaled air out through the mask vent holes. However, when the
device is not operating, insufficient fresh air will be provided through the mask, and
the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than
several minutes can, in some circumstances, lead to suffocation. This applies to
most models of CPAP devices.
•
In the event of power failure or machine malfunction, remove the mask.
•
The
S8 A
UTO
S
ET
V
ANTAGE
can be set to deliver pressures up to 20 cm H
2
O. In the
unlikely event of certain fault conditions, pressures up to 30 cm H
2
O are possible.
•
If oxygen is used with this device, the oxygen flow must be turned off when the
device is not operating.
Explanation:
When the CPAP device is not in operation and the oxygen flow is
left on, oxygen delivered into the air delivery tubing may accumulate within the
CPAP machine enclosure and create a risk of fire. This applies to most types of
CPAP machines.
•
Oxygen supports combustion. Oxygen should not be used while you are
smoking or in the presence of an open flame.
*
Ports may be incorporated into the mask or in connectors that are near the mask.
33887r2.book Page 6 Thursday, June 30, 2005 1:36 PM
Summary of Contents for S8 AutoSet Vantage
Page 6: ...E NGLISH 1 USER S MANUAL English S8 AutoSet Vantage System ...
Page 7: ...2 ...
Page 13: ...8 ...
Page 21: ...16 ...
Page 25: ...20 ...
Page 35: ...30 ...
Page 51: ...46 ...
Page 54: ...49 F RANÇAIS MANUEL UTILISATEUR Français S8 AutoSet Vantage ...
Page 55: ...50 ...
Page 69: ...64 ...
Page 73: ...68 ...
Page 81: ...76 ...
Page 85: ...80 ...
Page 95: ...90 ...
Page 103: ...98 ...
Page 106: ...101 E SPAÑOL MANUAL DEL USUARIO Español Sistema S8 AutoSet Vantage ...
Page 107: ...102 ...
Page 121: ...116 ...
Page 125: ...120 ...
Page 133: ...128 ...
Page 137: ...132 ...
Page 147: ...142 ...
Page 155: ...150 ...
Page 158: ...153 P ORTUGUÊS MANUAL DO UTILIZADOR Português Sistema S8 AutoSet Vantage ...
Page 159: ...154 ...
Page 173: ...168 ...
Page 177: ...172 ...
Page 185: ...180 ...
Page 189: ...184 ...
Page 199: ...194 ...
Page 207: ...202 ...