MWS_55_e.indb Page 94
geändert am 04.12.2020 von DO
Manual MWS 55 / MWS 88 / MWS 10 and Sensors 55
!Kx,D,TERM <Term><#13>
Generates a dependence of calculation of alias x from another alias input
by a term. If this term is true alias x and it's sensor is calculated, otherwise
output is -99999.9
<Term>: containing of
• Sensor prefix, or sensor name or $<sensor>
• <, <=, >, >= or =, <> (unequal) value or =, <> (range lower limit,
range upper limit)
i.e.
!K5,D,TERM WG > 0
; K5 only is calculated if WG > 0
!K2,D,TERM te = (-30,60)
; K2 only gets new values if TE is in the range of
-30...60
!Kx,D,A0|1[,a[,a..]]<#13>
;Disable / enable dependency
If the dependencyis disabled, sensor output runs normal, otherwise the
depen dent sensor postion is calculated with the term at first and if true the
value is updated.
Optionally attenuated real value, minimum, maximum, average and
standard deviation can be switched to the real sensor's value,
i.e.
!K5,D,A1,4,6
additionally swiches min and max.
8.2.1.3 Linearisation / calibration data
Sending linearization data:
!L<SENSORNUMBER>,<INDEX : 1..8>,F<RAW VALUE)><#13>
!L<SENSORNUMBER>,<INDEX : 1..8>,W <ANALOG VALUE><#13>
Detailed informations regarding the calibration of the MWS 55 you'll find on the "WetterStick":
../English/Manual/ Calibration_instructions_MWS.pdf
Setting the altitude for calculating the barometric pressure:
!O<ALTITUDE(m)><#13>
Setting the factor for correct measurent of the supply voltage (VCC):
!PL,G0.000504
Reset of precipitation measurement:
!R<X><#13>
X is the full hour for precipitation reset if storage interval is > 0 (logger on)
!R<#13>
Sets precipitation to 0 if storage interval = 0 (logger off)