55M4300CCE00 Rev A
2
CONFIGURATION
To ensure a successful, first calibration follow each calibration step in sequence, do not
miss out any steps unless they are marked as optional.
Many calibration screens show reminder notes or hints on the bottom line, remember
to read these for guidance.
Step #1: Accessing the calibration menu
1-
Press the Mode button.
2-
Scroll to highlight “Calibration Mode” using
“
”
or
“
”
.
3- Press
“
”
to enter this menu.
4-
The system will request a 5 digit password. The first digit will be highlighted, use
“
”
or
“
”
to select the first digit and press
“
”
to confirm and move on to the
second digit. Repeat this for all 5 digits, after confirming the 5
th
digit the system
will enter the calibration menu provided the password is correct. If the
password is not correct the system will return to the normal working screen.
5-
If a mistake is made while entering the password, press
“
”
to return to the
previous digit.
Some systems are installed with dual level password access, this manual includes
information on all features accessible from the top level. Some features will not be
available when the lower level password is entered and will not appear in the calibration
menu. It will not normally be necessary to access these features after the initial
installation is completed.
Provided the top level password is used the calibration menu will remain accessible until
the system is powered off otherwise the password will be required every time the menu
is selected.