QIAcube Connect Safety Instructions and Quick Start Guide 04/2019
15
Maintenance safety
WARNING/
CAUTION
Risk of personal injury and material damage
Only perform maintenance that is specifically described in this
user manual.
(W22)
WARNING
Risk of explosion
When cleaning the QIAcube Connect with alcohol-based
disinfectant, leave the QIAcube Connect hood open to allow
flammable vapors to disperse. Only clean the QIAcube Connect
when worktable components have cooled down.
(W23)
WARNING
Risk of fire
Do not allow cleaning fluid or decontamination agents to come
into contact with the electrical parts of the QIAcube Connect.
(W24)
WARNING
Risk of personal injury and material damage
To prevent the rotor nuts from loosening during operation of the
centrifuge, securely tighten the nuts using the rotor key supplied
with QIAcube Connect.
(W25)
WARNING
UV irradiation hazard
A mechanical lock ensures that the hood must be closed for
operation of the UV LED. If the hood sensor or lock is not
functioning correctly, contact QIAGEN Technical Services.
(W26)
WARNING
Risk of personal injury and material damage
Make sure that lids from spin columns and 1.5 ml
microcentrifuge tubes are in the correct position and pushed all
the way down to the bottom of the slots on the sides of the rotor
adapter. Incorrectly positioned lids can break off during
centrifugation.
(W27)