Connect Safety Instructions and Quick Start Guide 04/2019
WARNING
Moving parts
In case of breakdown caused by power failure, remove the
power cord and wait 10 minutes before attempting to
manually open the centrifuge lid.
(W19)
CAUTION
Damage to the instrument
After a power failure, do not move the z-module (robotic arm)
manually in front of the instrument. Damage may occur if the
QIAcube Connect hood is closed and collides with the z-
module.
(C8)
WARNING
Risk of personal injury and material damage
Raise the centrifuge lid carefully. The lid is heavy and may
cause injury if it falls.
(W20)
CAUTION
Risk of overheating
To ensure proper ventilation, maintain a minimum clearance of
10 cm at the sides and rear of the QIAcube Connect.
Slits and openings that ensure the ventilation of the QIAcube
Connect must not be covered.
(C9)
Heat hazard
The QIAcube Connect worktable contains a heated shaker.
WARNING
Hot surface
The shaker can reach temperatures of up to 70°C (158°F).
Avoid touching it when it is hot. Carefully remove the samples
after a run.
(W21)