PULSAR MEASUREMENT
87
To program the Ultra 3 for
Example 2 BS3680 U-Throated Flume
by using
the
Quick Setup
menu proceed as follows. If required access the
Program
Mode
, key in the passcode 1997 and press
ENTER.
Using the ‘right’ arrow
key, go to the
Quick Setup
menu press
ENTER
and as prompted, select the
relevant option and press
ENTER
.
QUESTION
OPTION
PMD Type
2 = 3680 Flume
3680 Flumes
3 = U-Throat
Calculation
1 = Absolute
No. of alarms
1 = 1 Alarm
Type alarm 1
2 = Low
Alarm No.1
1 = Set to relay 1
Transducer
9 = DUET
Volume units
2 = Cubic metres
Time Units
4 = Per hour
Measurement units
1 = Metres
Empty level
1.00m
Minimum head
0.00m
Maximum head
0.40m
Totaliser enabled
1 = On
Totaliser (R)
1 = Yes
Totaliser multiplier
7 = 1000
Approach dia. (Dim A) 0.70m
Throat dia. (Dim B)
0.50m
Throat length (Dim C) 1.00m
Programming is now complete, and the unit can now be returned to the run
mode, press
CANCEL
until
Run Mode?
Is displayed on the LCD press
ENTER
, and the Ultra 3 will return to
Run Mode
.
Important Notice
If relay setpoints do not meet the exact requirements of the application, they can
be modified to suit by pressing ENTER when, “For More Options Hit Enter”, is
displayed, and entering new values to relay setpoints as required. Alternatively,
access the setpoints by the main menu system or directly accessing the parameters.