IMPORTANT
: The resistance must be set to minimum to correctly perform
AutoCal.
l
Stop pedaling and wait for the pedals to completely stop. Wait 10 seconds to
allow the bike to complete the power measurement AutoCal function.
l
While continuously pedaling the bike at 60
RPM
, verify that the measured
workout
WATTS
at minimum and maximum resistance levels is within spe-
cification.
Resistance Level
Power (WATTS)
minimum
≤ 40 Watts @ 60 rpm
maximum
≥ 415 Watts @ 60 rpm
Power (WATTS) specification
IMPORTANT
: The bike must be AutoCal'd before making the WATTS meas-
urement. Also, the bike must be continuously pedaled between the min and
max resistance WATTS measurements. If the pedals stop, a new AutoCal may
occur invalidating the measurements.
Generator verification:
(Spinner
®
Chrono™ Power only).
l
Pedal the bike at 60
RPM
.
l
Verify that the measured generator voltage (service menu
GEN VOLT
) and cur-
rent output (service menu
GEN AMP
) is within specification.
Generator Spec
Power (WATTS)
Generator voltage (
GEN VOLT
)
approx: 9 Vac @ 60 rpm
Generator current (
GEN AMP
)
approx: 0.4 amps @ 60 rpm
Generator specification
Error Log &
ASL
1
verification:
(Spinner
®
Chrono™ Power only).
l
Verify that there are no current active error codes, view the service menu
ERROR LOG
. If there are logged error codes, resolve them and clear the error
log (service menu
CLEAR ERROR
l
Verify the ASL blue state (Some facilities may choose to switch the ASL func-
tion OFF, see
ASL SETUP
): Ride the bike and stop. Make sure that the console
ASL state is steady blue. If not, resolve any issues and reset the ASL to blue
state.
Return to service
(Spinner® Chrono™ Power models only)
1
Active Status Light: Service and maintenance status light.
Contact Precor Customer Support at [email protected] or 800.786.8404 with
any questions.
Page 15
3 Operation Verification
Operation Verification Checklist